General

The protection chamber for weighing OK is designed for safe handling of dangerous pharmaceutical substances, mainly for laboratory weighing of small quantities and preparation of samples for analyses. The suction chamber with horizontal, laminar filter-directed air flow prevents the passage of material from the chamber into the environment and thus protects the operator and the environment. It is a precise weighing, sampling, filtration or any operation with open, especially highly active substances.

Working principle

The suction chamber works on 100% air exchange (the amount entering also exits the system - without recirculation). The chamber takes air from the surrounding area - the laboratory (from the front through the working aperture).

The air enters the workplace of the suction chamber from the front through the working aperture, using a fan mounted in the suction unit. The captured air then travels over the workplace and the first filtration quality H14 (EN 1822). Through outlet filters, the incoming air, which is contaminated with the substance within the working place, filters and purifies. The filtered air then returns to the place.

The first output HEPA filter quality H14 is installed directly behind the workplace. It is safe to replace the first output filters that are loaded with dangerous particles. The second output HEPA filter quality H14 is integrated into the suction unit and provides additional protection and assumes the role of air filtration in the event of a change or damage to the first or the other. the first two output filters. Through outlet filters, the incoming air, which is contaminated with the substance within the workplace, filters and purifies. The filtered air then returns to the place.

OK Air flow

Control

The user screen is for visualization and more transparent management. It is located on the front of the chamber, next to  the user. Its basic task is to manage the system, display errors and manage service settings.

 

Specifications

  • All surfaces and corners built according to GMP, GLP standard
  • Metal construction surface protection with powder coating in color RAL 9010
  • Granite working surface, thickness 50 mm
  • Working place surrounded by acrylic glass (PMMA)
  • Acrylic glass lift door, for the inlet of larger equipment
  • Good illumination of workplace
  • Noise <60 dB
  • With a built-in fan and the possibility of adjusting the intensity of the suction
  • Microprocessor control of operation and safety functions (alarming of limit values)
  • User-selection display mode and parameter setting (in case of validation and service)
  • Double HEPA filtration
  • Integrated connectors for performing the integrity of absolute H14 filters (DOP / DEHS test)

Options

  • Stainless version
  • A drainage module without a fan, with an outlet for extracting air from the room
  • Faucet for gas, air, vacuum
  • Elevated work spac
  • Dome, to increase work space
  • Movable shield with gloves for increased safety
  • Built-in charcoal filter
  • Outlet port for material or waste
  • CLS for the outlet of material
  • Welding tongs
  • El. socket
  • Performance according to customer requirements

Type

OK 9

OK 12

OK 15

OK 18

Outside dimen. with stand
W x D x H [mm]

940x866x1476

1240x906x1476

1540x906x1476

1840x906x1476

Outside dimen. without stand
W x D x H [mm]

940x866x746

1240x906x746

1540x906x746

1840x906x746

Working room
W x D x H [mm]

843x576x481

1143x576x481

1443x576x481

1743x576x481

Size of the working opening
W x D x H [mm]

843x182,5

1143x182,5

1443x182,5

1743x182,5

Size of the suction unit
W x D x H [mm]

330x605x664

550x675x664

550x675x664

550x675x664

Max. connecting power [W]

750

750

750

750

Air flow velocity [m/s]

> 0,40

> 0,40

> 0,40

> 0,40

Operating voltage [V/Hz]

 230/50

 230/50

 230/50

 230/50

Environment (working)

Temp.: 10-30 °C

RH: up to 70%

Temp.: 10-30 °C

RH: up to 70%

Temp.: 10-30 °C

RH: up to 70%

Temp.: 10-30 °C

RH: up to 70%

Weight [kg]

210

290

320

360

ABS. Filter H14
[dim. mm/pcs]

457x610x69 / 1

457x457x69 / 2

457x610x69 / 2

457x762x69 / 2

ABS. Filter H14
Ø 200 x 500mm [pcs]

1 2 2 2

 