Safety cabinet for preparation of radiopharmaceuticals MC Pb
Protection of the operator, the product and the environment while working with radioactive isotopes.
General
MC Pb Safety cabinet class II assures the highest level of safety for the operator, offers clean conditions and protects the environment from contamination. The device fulfills :
- Safety class II - SIST EN 12469:2000
- Class II, type A2 – NSF:2008
- Working area GMP class A by GMP or- class 5 by ISO 14644-1:1999
It is constructed and tested in accordance with SIST EN 12469: 2000.
Advantages
- Easy for cleaning and maintenance
- Quick and safe filter change
- Easy disinfection
- Low noise ventilator
- Front glass panel adjustable in height
- Progressively adjustable ventilator
- Available in different sizes
- Easy accessible electric section
Working principle
The chamber takes part of the air from the room and returns it to the room through an absolute outlet filter, the rest of it circulates inside the chamber.
The air enters the chamber from the front side at the lower part of the work space through the front opening and goes on through the cuttings in the segments of the work desk. Below the segments of the work desk the inlet and recirculation air are mixed.
Further on, the air travels through the return chanel to the upper housing and enters the over pressure hood. The fan pushes part of the air (approx. 30%) through the output filter H14 (EN 1822) to the surrounding area, while the remaining part of the air (ca.70%) is pushed into the working space through the filter above the workplace of quality H14 (EN 1822) and distribution grids. Output / inlet and recirculation air is provided by the ratio of the exit filter and filter surface area over the working area.
The distribution grid laminarally distributes air over the working area and vertically directs it towards the working surface of the chamber. The laminar air flow takes away dust particles generated by material manipulation. An air curtain is formed on the front of the chamber, between the operator and the location of the dust, which separates the working space in the chamber from the surrounding area.
The air curtain protects the work area from possible external contamination and prevents airborne particles from flowing out of the work space.
In addition to the described environment protection and the operator through air filtration, the MC Pb provides the protection of the operator against radiation while using radiopharmaceuticals. For this purpose, the chamber is coated with lead, thickness 10 to 80 mm.
Control
A visual touch screen is designed for visualization and more transparent management. It is located on the front of the chamber, next to the user. Its basic task is system management, error display and service settings.
Specifications
- Surface-protected metal construction with powder coating in color RAL 9010
- Workplace and work segments made of high quality stainless steel sheet
- The workplace is entirely made of brushed stainless steel AISI 304 (EN 1.4301)
- Work segments are made of brushed stainless steel AISI 316L (EN 1.4404)
- Work bench consists of segments, easy to clean and autoclave
- Height of working surface for half sitting-position - 950 mm
- All surfaces and corners built according to GMP, GLP standard
- Construction of chamber constructed under a slope of 8 ° - easier sitting and working of the operator
- Good illumination of workplace (> 1000 lux)
- Noise <60dB
- Lateral visibility of the workplace, sides of safety glass
- Motorized raising of the front safety glass (automatic layout to the appropriate height for the selected mode)
- Microprocessor control of operation and safety functions (alarming of limit values),
- User-selection display mode and parameter setting (in case of validation and service)
- Absolute filtration of exhaust air and air in the field of work, with quality filters H14-EN1822: 2010
- Integrated connectors for performing the integrity of absolute H14 filters (DOP / DEHS test)
- Connections in the field of work:
- 2x socket 220V / 2,5A
- 2x USB connector (connection internal workplace - external side)
- Work equipment for the preparation of radiopharmaceuticals:
- Protection of the workplace 10 mm Pb
- Protective movable glass equivalent to 10 mm Pb
- Drawer for two generators 50 mm Pb
- Waste drawer 10 mm Pb
- Drawer for activity meter 20 mm Pb
Options
- Stainless version
- Built-in UV bactericidal bulb
- In the back wall there is a built-in monitor
- Suction hood
- Faucet for gas, air, vacuum
- El. sockets, USB, RJ-45 connections
- Isokinetic probe
- Speed sensor
- Temperature and humidity sensor
- Work equipment for the preparation of:
- Protection of the workplace 10, 30, 50 mm Pb
- Additional front shield height 150 mm 10, 30, 50 mm Pb
- Protective slide glass equivalent to 6.1, 10, 20, 30 mm Pb
- Drawer for two generators 50, 80 mm Pb
- Waste drawer 10, 30 mm Pb
- Drawer for activity meter 20, 50 mm Pb
- Performance according to customer requirements
Type |
MC 12 Pb 10 |
MC 15 Pb 10 |
MC 18 Pb 10 |
Outside dimension |
1340x900x2300 |
1650x900x2300 |
1960x900x2300 |
Workplace |
1190x600x600 |
1500x600x600 |
1810x600x600 |
Max. connection power without sockets [W] |
1000 |
1500 |
1500 |
Max. connection power on socket [W] |
550 |
550 |
550 |
Laminar air flow [m/s] |
0,4 |
0,4 |
0,4 |
Voltage [V/Hz] |
230/50 |
230/50 |
230/50 |
Weight [kg] |
1900 |
2200 |
2400 |
ABS. working filters H14 |
305x610x69/4 |
305x610x69/5 |
305x610x69/6 |
ABS. exhaust filters H14 |
305x457x69/2 |
305x457x78/2 |
305x457x69/3 |