Laminar horizontal chamber LFH
The laminar horizontal chamber of the LFH series ensures a high level of protection for the product - ISO 5 (ISO 14644-1), class A (GMP).
General
The laminar horizontal chamber, with its action, allows the product-substance to be protected from contamination with particles from the environment. Clean working conditions are ensured within the workplace. Filtered airflow (HEPA filtration H14), which moves horizontally from the back wall of the work space to the operator, prevents particulate inlet and impurities from the environment into the workplace. The filtered airflow is arranged laminarally and homogeneously over the entire surface of the workplace.
Advantages
- Easy maintenance
- Fast and safe filter change
- Built-in low-noise fan
- Progressively adjustable fan
- Availability in various dimensions
- An easily accessible electrical part
- Easy management
Working principle
The chamber sucks the air from the environment through an input pre-filter G4 quality that is mounted into the upper part of the chamber. Then, the inlet fan pushes the air through an absolute operating filter H14 (EN 1822) mounted into the rear part of the chamber. The distribution grid arranges the air across the entire working surface in a horizontal one-way stream. The air then travels through the openings to the front of the chamber. One-way airflow carries dust particles from the processing of materials in the workplace. The control system consists of an internal control unit and a control panel, from which the user can operate the chamber. The control unit controls the fan and provides:
- Constant frequency control of motors due to clogging of filters
- Balanced air flow within the safety chamber
- Indicator of possible damage or an alarming state of the system
Control
User handling with the device is done by using a control panel, through which the user can choose between different laminar airflow speeds, light / turn off lights and control various optional components.
Specifications
- Surface of metal construction protected with powder coating in RAL 9010
- Work area and working segments made of quality stainless steel
- Working surface made of segments what enables easy cleaning and autoclaving
- All surfaces and corners built in accordance with GMP, GLP standard
- Good illumination of work area (> 500 lux)
- Noise < 60dB
- Side walls with integrated safety glass
- Operating and safety features controlled by microprocessor (alarm limit values)
- Different modes and setup of working parameters selected with user interface (in case of validation and service)
- Integrated connector for absolute filters H14 integrity test (DOP/DEHS test)
Options
- Stainless version
- Faucet for gas, air, vacuum
- El. socket
- Performance according to customer requirements
Type |
LFH 9 |
LFH 12 |
LFH 15 |
LFH 18 |
Outside dimension |
1020x900x1940 |
1325x900x1940 |
1630x900x1940 |
1935x900x1940 |
Workplace |
947x550x715 |
1252x550x715 |
1557x550x715 |
1862x550x715 |
Max. connection power [W] |
2050 |
2050 |
2050 |
2100 |
Max. power on socket [W] |
550 |
550 |
550 |
550 |
Laminar air flow [m/s] |
0,25÷0,45 |
0,25÷0,45 |
0,25÷0,45 |
0,25÷0,45 |
Voltage [V/Hz] |
230/50 |
230/50 |
230/50 |
230/50 |
Weight [kg] |
190 |
210 |
230 |
250 |
ABS. working filter H14 |
910x610x69 |
1220x610x69 |
1524x610x69 |
1830x610x69 |
Inlet pre-filter |
G4 / EN779 |
G4 / EN779 |
G4 / EN779 |
G4 / EN779 |